Key Product Features & Technical Specifications

Key Product Features

   PC or notebook computer software and transmission compatibility
   Portable, lightweight unit with „slip on“ belt system for chest or back mounting
   Measurement programs: CPET, Spirometry/Flow-Volume , ECG (optional), Interpretation Help
   Long-range telemetry with real-time monitoring, and/or internal data storage
   Complies with ERS and ATS guidelines

Oxycon Mobile


Technical Specifications

 Oxycon Mobile unit
Method: Breath by breath
Dimensions (LxWxH): 2 units, 126 x 96 x 41 mm each
Weight: 950 g including belt, battery and mask
Data memory: 32 MB flash card (5 hours recording capacity)

 Volume and gas measurement
 
Ventilation (V’E):
O2 uptake (V’O2 ):
CO2 release (V’CO2 ):
RER:
Range Accuracy
0 to 300 l/min 2% or 0.05 l/min
0 to 7 l/min 3% or 0.05 l/min
0 to 7 l/min 3% or 0.05 l/min
0.6 to 2.0 4%
Volume sensor  
Type: TripleV
Volume: 0 to 10 l
Accuracy: 50 ml or 2%
Resolution: 3 ml
Flow: 0 to 15 l/s
Accuracy: 70 ml/s or 3%
Resistance: <0.1 kPa/l/s at 15 l/s
O2-Analyser  
Principle: electrochemical
Range: 0 - 25%
t90: 80 ms (after filtering)
Accuracy: 0.05 vol%
CO2-Analyser  
Principle: thermal conductivity
Range: 0 to 10%
t90: 80 ms (after filtering)
Accuracy: 0.05 vol%

 Ambient Specifications
Temperature: -10°C to +50°C
Pressure: 500 - 1100 hPa
Altitude: approx. -1.400 to 5.500 m
Humidity: 10 - 95%

 Battery
Type: Lithium-Ion battery
Capacity: 1500 mAh
Life: 2 hours
Voltage: 3.6 V
Weight: 45 g

 Telemetry
Type: Spread spectrum, frequency hopping, full duplex
Power: 100 mW
Frequency: 2401 - 2495 MHz
Range: up to 1.000 m

 Heart Rate/ECG
Heart rate: Polar belt
Heart rate (option): 1 channel, 3 leads
ECG (option): 8 channels, 12 leads
Leads: Einthoven, Wilson, Nehb, Goldberger
Frequency: 0.05 -100 Hz
Gain: 5, 10, 20 mm/mV
CMRR: >120 dB
Sampling rate: 240 Hz or 500 Hz, 3 or 8 channels
A/D resolution: 12 bit