Compact CPX system for precise and reliable determination of cardiorespiratory and resulting parameters for all groups of patients. Breath-by-Breath and Intrabreath measurement.
The optional 12-channel ECG is completely integrated in the system.
Technical Data
Hardware and software
Compact housing including O2 and CO2 analyzer
12-channel ECG (option)
Ventilation measurement based on a highly-precise digital volume sensor (TripleV sensor)
Automatic calibration programs
Pentium computer
Ink-jet printer
MS-Windows XP
Interfaces for connecting peripheral equipment
TripleV volume sensor
Range
0 to 10 L
Accuracy
50 ml or 2 %
Resolution
3 ml
Flow
0 to15 L/s
Accuracy
0.07 L/s or 3 %
Resistance
<0.1 kPa/L/s at 15 L/s
Oxygen analyzer
Range
0 to 25 %
Accuracy
0.05 %
Carbon dioxide analyzer
Range
0 to 10 %
Accuracy
0.05 % (0 to 5 % CO2)
ECG
Channels
12
A/D resolution
12 bits
Frequency range
0.05 to 100 Hz
Gain
5, 10, 20 mm/mV
CMRR
> 120 dB
Leads
Einthoven, Wilson, Nehb, Goldberger
Dimensions
Height x Width x Depth
180 x 70 x 90 cm
Heart rate monitoring with Polar belt (option)
Application
Heart rate measurements, only
Battery
Integrated lithium battery; approx. 2500 hours
Materials
Buckle: Polyurethane
Tissue: Nylon, polyester and natural rubber latex